[ViZion DR+ 1417V2 Wireless Panel Installation Manual]
[Ultra 4.2.9.0]
24
Customer Support
1.800.366.5343
–
4.7.4 Panel Calibration Setup
1. Select the
Acquire
tab in iDetector.
2. Uncheck the
Defect
,
Gain
, and
Offset
checkboxes
in that order
.
Note
: during the calibration, please do not
move the detector once calibration
has been started and keep Link
cable connected to the detector.
3. Select the
Calibrate
tab, check the
radio button for
PostOffset
, and click
the
Start Generate Templates
button.
The Calibration wizard will open into a new
window, presenting the
Welcome
screen.
4. Set X-ray environment:
•
Tube SID
=
50” or higher (at least
greater than 47”).
•
No Collimation
= open collimator
blades at least 1” past the edges of
the panel on all sides.
•
No Grid
= remove any grid or
pressure grid cover.
•
No Objects
= remove any objects
from the surface of the panel.
5. Click the arrow in the top right corner
labeled
Start the journey of
Calibration
.