[ViZion DR+ 1417V2 Wireless Panel Installation Manual]
[Ultra 4.2.9.0]
26
Customer Support
1.800.366.5343
–
11. After taking all 5 gain exposures, click
the
Back
button if you want to retake
any previous exposures, then click
Prep
and retake the exposure. Otherwise, if
you are satisfied with all exposures,
click the
Next
button in the top right
corner.
12. Wait for the
Gain Map Generation
to
complete and then click
Next
.
4.7.6 Panel Defect Calibration
Note
: the Defect Calibration screen
requires 19 exposures. The left-
hand side of the screen tells you the
suggested kV and mAs to use for
the current exposure. The right-
hand side of the screen is divided
into 19 blank defect exposure frame
icons. Each exposure frame is
numbered sequentially, with the kV
to use noted next to each one. A
green box surrounds the current
exposure frame.
13. Set your techniques to
70kV / 1.1 mAs
and take your
first
exposure.
Note
: after taking the first exposure, the
current defect exposure frame will
still be surrounded in green. It will
either have a green checkmark, if it
was successful, or a yellow
exclamation point, if it was not
within 100 +/- of the expected
center average of 1000.