4.6 Device Pool Configuration
After successful driver pack installation, it is
time to configure the device settings for the
ViZion DR+ 1417V2 wireless panel.
Before performing the following steps, make
sure the wireless panel is powered on, with
a fully charged battery, and connected with
the Link cable.
1. After closing the Device Manager
window you should be back at the
Options menu with the
Device
Configurations
tab selected.
2. Click to highlight
DEFAULT
under the
“Configuration Pool” section.
3. Click on
ViZion DR+
and the panel
device configuration section will appear
below it.
4. Scroll down to the
Device
section and
click the button with
3 dots
, to the right
of “S/N” field. This will perform a scan
and automatically populate the S/N and
Path fields.
5.
By default “Check Shock Log” is
unchecked and “Manage Automatically”
is checked.
6. Select the drop down next to
“Parameters Template” and choose
Mars1417V
.
7. Finally, click on the
Utility
button to
open the iDetector interface.
Note
: after clicking on Utility, the Maven
software will shutdown
communication between the panel
and the Ultra UAI software and
close, and then iDetector software
will launch automatically as a popup
window and begin communication
with the panel. Settings inside the
panel can then be changed or panel
calibrated.