[ViZion DR+ 1417V2 Wireless Panel Installation Manual]
[Ultra 4.2.9.0]
11
Customer Support
1.800.366.5343
–
2.2.2 SourceRay SR-130 Accessory Cables
1. Connect the hand switch cable to the
Generator Console (or directly to the
tube head, depending on model X-ray
unit being used).
2. The power cable connects the X-ray unit
to a grounded wall outlet.
3 Network Installation
3.1 Fixed Room Network
Configuration
After installing the hardware, configure the
network address on the
Intel Gigabit CT PCI-e
network adapter
(the add-on network card on
the acquisition PC).
Note
: the ViZion DR+ Wireless panel,
acquisition PC, and ASUS access
point, all come preconfigured.
These instructions show how to
configure the devices, as needed.
The goal is to ensure that each device can
communicate with one another, which requires
that they are configured on the same network
segment.
The following network equipment is required for
fixed room environments:
•
ASUS Dual-Band Wi-Fi Gigabit AP
•
75’ Cat5 Ethernet Cable
•
Link Cable (to preconfigure the ViZion
DR+ 1417V2 wireless panel)
You must first configure the IP address on the
acquisition PC.
By default, the ViZion DR+ 1417V2 wireless
panel will have an IP address hardcoded to
192.168.8.8
. The ASUS AP should be
configured with an IP address of
192.168.8.50
.
The CT network adapter on the PC should be
configured with an IP address of
192.168.8.188
.
The following diagram illustrates a fixed room
network environment.