4.4 Device Licensing
For panel, generator or positioner device
licensing, contact Konica Minolta DR Support.
4.5 Device Driver Installation
The ViZion DR+ 1417V2 wireless panel uses a
different SDK version than 1417V panels, and it
is only compatible with
Ultra UAI versions
4.2.9.0 or higher
, which introduces a selection
for Third Generation Venu drivers.
Note
: the Venu Drivers Generation
selection setting is located in the
Developer tab under Options menu.
ViZion DR+ 1417P, 1417P1, 1717M
and 1717M2 panels should be
configured to use First Generation
Venu drivers.
ViZion DR+ 1417V and 1717M2.5
panels should be configured to use
Second Generation Venu drivers.
ViZion DR+ 1417V2 and 1012V2
panels should be configured to use
Third Generation Venu drivers.
1. Before installing the new venu driver
pack, double-click on the
Konica
Minolta logo
, located in the bottom left
corner of the
Options
menu. This action
will display the Developer tab in the
upper right corner.
2. Select the
Developer
tab and select the
drop down next to
Venu Drivers
Generation
and select
Third
.
3. Click
Apply
, and then click
OK
to the
message that pops up.
4. Click
OK
to exit Option menu, and then
click the
Log Off
icon in the top right
corner to exit Ultra.
5. Confirm by clicking the
Log Off
button.
4. Double-click on the
Ultra
icon to
relaunch the software.