34
Recording the Measured Illuminance Continuously (Analog Output)
1
Have the instrument ready for operation as described in
“Measuring the illuminance” (page 13) and push out the hold
button (RUN).
z
The range can be switched in RUN state only.
2
Insert the Mini plug (optional) into the analog output terminal.
z
How to connect the Mini plug (optional)
When connecting the instrument to an external device, solder the Mini plug
as shown in the figure. Be sure to connect a shielded cable to the Mini plug.
3
Select the measuring range suitable for the illuminance to be
measured.
z
Refer to “Setting the Range Manually” (page 32).
4
Adjust the sensitivity of the recorder according to the
illuminance to be measured.
z
The analog output level of this instrument is 1 mv/digit (one display
count), and the maximum output voltage is 3 V (3,000 mV).
5
Attach the cap to the receptor head (or block the head to
prevent entry of external light).
“0 lx” will appear in the display section. However, the analog
output voltage may not be 0 mV due to the offset voltage.