Digital Output of the Measured Value
This section explains how to print out the measured data.
To print out the data, the printer cable T-A12 (optional) is required to connect a printer to the instrument.
Printer Requirements
The printer must meet the following requirements.
Number of print digits
: 27 or more
Data input
: RS-232C
Data control
: Busy
Baud rate
: 9,600
Character length
: 7 bits
Parity
: Even
Stop bit
: 1 bit
Basic function
: CR (0DH) carriage return
Example of recommended printer
Model name: DPU-H245AS-A03A (Seiko Instruments Inc.)
Procedure
Connect the instrument to the printer with the printer cable T-A12.
z
Before connecting them, make sure that the instrument’s power switch is set to “O” (OFF).
z
Always set the instrument’s power switch to “I” (ON) before turning ON the printer’s power switch. If the
printer’s power switch is turned ON first, the printer may not work properly.
1
Make sure that the instrument’s power switch is set to “O”
(OFF), and connect the instrument to the printer with the
printer cable TA12.
z
The illustration on the left shows an example of
connecting the instrument to the printer.
▼
Connection to the printer
Printing the Measured Data (Digital Output)
DIP switch settings
Содержание T-10A
Страница 1: ...ILLUMINANCE METER T 10A T 10MA INSTRUCTION MANUAL Before using this instrument please read this manual...
Страница 9: ...7 Basic Operations...
Страница 28: ...26...
Страница 29: ...27 Advanced Operations...
Страница 47: ...45 Accessories...
Страница 51: ...49 Explanation...
Страница 60: ...9222 A58U 11 BHJFKK Printed in Japan 2012 KONICA MINOLTA INC...