50
Error Messages
An error may occur during operation due to some problems. If an error
occurs, refer to the table below and take the necessary action.
Message
Description
Action
E r r U
E r r E
Initial communication error No
response from the receptor head
during initial communication
Check connection. If the instrument is
connected properly, contact the nearest
Konica Minolta-authorized service facility.
Start command error The receptor head
has not received the start command,
because, for instance, the receptor head
was not connected to the main body at
the time the power was turned ON.
Set the power switch to “I” (ON)
again. If this message reappears,
contact the nearest Konica Minolta-
authorized service facility.
E r r
EEPROM error Error with the EEPROM
data (in the receptor head)
Remove the batteries (or AC adapter),
then re-install them. If this message
reappears, contact the nearest Konica
Minolta-authorized service facility.
E r r A
Preset value error The value is
outside the settable range.
Check the settable range, and make
sure that the value is inside the range.
E r r C
Communication error (main body
↔
receptor head) Communication
between the main body and receptor
head was not performed properly.
Check connection. If the instrument is
connected properly, contact the nearest
Konica Minolta-authorized service facility.
Содержание T-10A
Страница 1: ...ILLUMINANCE METER T 10A T 10MA INSTRUCTION MANUAL Before using this instrument please read this manual...
Страница 9: ...7 Basic Operations...
Страница 28: ...26...
Страница 29: ...27 Advanced Operations...
Страница 47: ...45 Accessories...
Страница 51: ...49 Explanation...
Страница 60: ...9222 A58U 11 BHJFKK Printed in Japan 2012 KONICA MINOLTA INC...