52
[Conforms to JIS]
Ideally, the relative spectral responsivity of the illuminance meter should match V (
λ
)
the spectral luminous efficiency of the human eye for photopic vision.
As shown in the graph above, the relative spectral responsivity of Konica Minolta Illuminance
Meters T-10A/T-10MA is within 6% (f
1
’
) of the JIS spectral luminous efficiency V (
λ
).
Class
General class Precise
General class AA
General class A
f
1
’
(%)
3
6
9
JIS ; Japanese Industrial Standards
f’1 (JIS’s symbol) ; The degree to which the relative spectral responsivity matches
V (
λ
) is characterized by means of the error f
1
’
.
Note : The contant has the dimension nm
-1
S’(
λ
)
rel
normalized relative spectral responsivity
P(
λ
)
A
Spectral distribution of the illuminant used in the calibration
(standard illuminant A)
S(
λ
)
rel
Relative spectral responsivity with arbitrary reference
V(
λ
)
Spectral luminous efficiency of the human eye for photopic vision
d
λ
Wavelength pitch
λ
1
Lower limit of visible wavelength region
λ
2
Upper limit of visible wavelength region
S ( )
rel
=
P(
λ
)
A
V(
λ
)
d
λ
P(
λ
)
A
S(
λ
)
rel
d
λ
•
S(
λ
)
rel
λ
λ
1
λ
2
ʃ
λ
1
λ
2
ʃ
′
0
10
20
30
40
50
60
70
80
90
100
Relativ
e spectr
al response (%)
750
700
650
600
550
500
450
400
Wavelength (nm)
The spectral luminous efficiency
Konica Minolta Illuminance Meters T-10A/T-10MA
f
1
′
=
λ
1
λ
2
S
′
(
λ
)
rel
-V(
λ
)
d
λ
V(
λ
)
d
λ
×
100(%)
ʃ
λ
1
λ
2
ʃ
Relative Spectral Response
Содержание T-10A
Страница 1: ...ILLUMINANCE METER T 10A T 10MA INSTRUCTION MANUAL Before using this instrument please read this manual...
Страница 9: ...7 Basic Operations...
Страница 28: ...26...
Страница 29: ...27 Advanced Operations...
Страница 47: ...45 Accessories...
Страница 51: ...49 Explanation...
Страница 60: ...9222 A58U 11 BHJFKK Printed in Japan 2012 KONICA MINOLTA INC...