31
4
Press the [SET] key.
The instrument is now ready for you to enter a value, and
the currently displayed measured value will blink.
5
Enter the value using the [
▲
] and [
▼
] keys.
Each time the [
▲
] or [
▼
] key is pressed, the value will change in steps of
the smallest digit.
Holding the [
▲
] or [
▼
] key will cause the
displayed value to change continuously.
z
To exit setting, press the [CALL] key.
6
When you have entered the value, press the [SET] key.
The CCF will be calculated from the entered reference value and then set.
z
If the calculated correction factor is not within the range from
0.500 to 2.000, the display shows "Err A", and then returns
to the beginning of the SET mode. Reenter the values.
z
To check the currently set value, press the [CALL] key. The currently set value will be displayed while
the [CALL] key is held down.
Содержание T-10A
Страница 1: ...ILLUMINANCE METER T 10A T 10MA INSTRUCTION MANUAL Before using this instrument please read this manual...
Страница 9: ...7 Basic Operations...
Страница 28: ...26...
Страница 29: ...27 Advanced Operations...
Страница 47: ...45 Accessories...
Страница 51: ...49 Explanation...
Страница 60: ...9222 A58U 11 BHJFKK Printed in Japan 2012 KONICA MINOLTA INC...