394149/D
259
b
Allow the data recording to run approximately five minutes.
c
Stop raw data recording.
d
Use a file manager, and verify that the recorded file(s) have been saved on
the chosen disk.
e
Verify that the playback is operational.
14
Select
Help
on the top bar.
a
Make sure that the online help opens on its start page.
b
Close the online help.
Result
Requirements
Results
The latest software version is installed.
Each transducer is installed with the correct parameters.
Each transceiver channel is operational.
The EK80 detects and displays a bottom echo in each echogram.
Date and signature:
Related topics
Measuring noise in passive operating mode
Low noise is a key factor for high quality and reliable measurements. The performance
of the EK80 will always be limited by different noise sources. The noise is measured
while the EK80 operates in
Passive
mode with the transmit pulses disabled.
Prerequisites
The EK80 hardware units are installed as specified in this manual.
• The EK80 system is turned on and operates normally.
Caution
You must never set the EK80 to "ping" unless the transducer is submerged in water.
The transducer may be damaged if it transmits in open air.
• All the relevant transceivers have been set up, and they are operational with their
respective transducers.
• All relevant channels (transceiver/transducer combinations) are installed in the user
interface.
• All relevant external sensors are connected to the EK80. The sensors are turned
on and operate normally.
Содержание Simrad EK80
Страница 1: ...kongsberg com simrad Simrad EK80 INSTALLATION MANUAL ...
Страница 2: ......
Страница 12: ...10 394149 D Simrad EK80 ...
Страница 223: ...394149 D 221 6 Start the installation by selecting Next 7 Select Finish to close the wizard Setting to work ...
Страница 313: ...394149 D 311 Drawing file ...
Страница 314: ...312 394149 D Simrad EK80 Installation Manual ...
Страница 317: ...394149 D 315 Drawing file ...
Страница 318: ...316 394149 D Simrad EK80 Installation Manual ...
Страница 320: ...318 394149 D 201575 Transducer connector assembly and wiring Simrad EK80 Installation Manual ...
Страница 352: ... 2020 Kongsberg Maritime ISBN 978 82 8066 182 1 ...
Страница 353: ......
Страница 354: ...Installation Manual Simrad EK80 ...