SensoGate WA131M
37
6.3.3 Drive Unit: Assembly
Note:
The radial installation position of the drive unit is determined by a coding pin in the calibration
chamber and a hole in the drive unit. The coupling nut can be tightened only if the drive unit is cor-
rectly inserted into the process unit.
1
2
3
4
5
6
7
8
9
01.
Move the drive unit to the SERVICE position.
Moving into the SERVICE Position, p. 26
02.
Push the driv
e unit
(1)
with the immersion tube
(3)
into the proc
ess unit
(5)
. While doing so, posi-
tion the c
oding pin
(4)
in the hole
(2)
.
Note:
Do not tilt the coupling nut. Use a suitable spanning wrench (e.g., the one contained in
ZU0680 ser
vic
e set or ZU0740 ser
vice set).
03.
Position the c
oupling nut
(8)
and tighten clockwise finger tight or t
o 10 N
m using the spanning
wr
ench
(6)
.
04.
As required, install the outlet hose at the outlet
(7)
.
Outlet Hose: Installation, p. 23
05.
Optional: Install the inlet hose
1)
.
Inlet Hose: Installation, p. 23
06.
Optional: Install the limit switch
1)
.
07.
As required, install the sensor
(9)
.
Installing and Removing Sensors, p. 27
1)
Availability is dependent on the ordered version.