SensoGate WA131M
34
6.1.4 Immersion Lock Without a Mounted Liquid-Electrolyte Sensor: Functional Test
Note:
The safeguard “Immersion lock without a mounted liquid-electrolyte sensor” can be seen at the
yellow indicator ring
(2)
above the bellow
s
(1)
.
1
2
3
4
01.
Move the SensoGate WA131M into the SERVICE position.
Moving into the SERVICE Position, p. 26
02.
Loosen the small c
oupling nut
(4)
a little; do not loosen completely.
WARNING! In the event of a malfunction, pressurized process medium may escape from the
SensoGate WA131M.
Do not completely loosen the large coupling nut
(3)
to ensure that pressure
resistance is still available in the event of a malfunction.
03.
Loosen the large c
oupling nut
(3)
around 1.5 rotations; do not loosen completely.
04.
Check the function of the “Immersion lock without a mounted liquid-electrolyte sensor”.
✓
It must be impossible to depress the safety lock button.
✓
It must be impossible to rotate the rotating collar.
05.
Fasten the large c
oupling nut
(3)
finger tight.
06.
Fasten the small c
oupling nut
(4)
finger tight.
07.
Move the SensoGate WA131M into the PROCESS position.
Moving into the PROCESS Position, p. 25
✓
Safety lock button pops out when the PROCESS position is reached.
✓
Rotating collar is locked to prevent rotation.
08.
Repeat the functional test every 12 months. As applicable, adjust the interval to match the specific
application for which the SensoGate WA131M is used.