SensoGate WA131M
22
3 Installation
3.1 Retractable Fitting: Installation
WARNING! Risk of explosion from mechanically generated sparks when used in explosive at-
mospheres.
Take measures to prevent sparking. Follow the safety instructions.
Operation in Explosive Atmospheres, p. 9
Note:
The possible installation angle is 10° above the horizontal plane. An installation angle of 360°
(i.e., upside down) is only permitted if using sensors approved for upside-down installation.
10°
10°
01.
Check scope of delivery of the SensoGate WA131M for completeness.
02.
Check the SensoGate WA131M for damage.
03.
Ensure the required sensor installation clearances.
04.
Fasten the SensoGate WA131M to the process port using the process connection.
05.
Optional: If using the product in explosive atmospheres, connect the grounding connection of
SensoGate WA131M to the plant’s equipotential bonding system.
See also
Operation in Explosive Atmospheres, p. 9
3.2 Safety Accessories: Installation
Consult the related instructions for information on installing the safety accessories (e.g., ZU0818
retainer clamp).
See also