6.3.7 Calibration Chamber: Installation
Note:
ZU0754 ser
vic
e set or ZU0740 ser
vice set are required to install the calibration chamber.
Note:
To ensure correct assembly of the O-rings and the scraper ring, use the accessory tools ZU0746
and ZU0747. The procedure for handling the accessory tools is described in the relevant documenta-
tion.
1
2
3
4
01.
Check the O-rings and scraper ring for damage; replace the O-rings and scraper ring if necessary.
02.
Connect the top
(1)
and the bottom
(2)
of the calibration chamber and screw together finger
tight.
03.
Position the pliers
(4)
and use the face pin spanner wrench
(3)
to screw the calibration chamber
together.
5
5
Note:
Securing the calibration chamber with the two screws is not possible until the top and bottom
parts have been firmly screwed together (to the hard stop).
04.
Tighten the screw
s
(5)
with a screwdriver of type T
X25 .
6.3.8 Knick Repair Service
The Knick Repair Service offers professional corrective maintenance on the SensoGate WA131M to the
original quality. Upon request, a replacement unit can be obtained for the period of the repair.
Further information can be found at www.knick.de.