6.3 Corrective Maintenance
6.3.1 Corrective Maintenance Safety Instructions
WARNING! Process medium, possibly containing hazardous substances, can escape from the
SensoGate WA131M.
Follow the safety instructions.
CAUTION! Risk of cutting injuries from broken sensor glass.
Handle the sensor with care. Follow
the safety instructions in the related sensor documentation.
6.3.2 Drive Unit: Disassembly
1
2
3
4
5
01.
Safely disconnect the SensoGate WA131M from the process.
Retractable Fitting: Removal, p. 44
02.
As necessary, disconnect the outlet hose, inlet hose
1)
, and limit switch
1)
.
03.
Move the SensoGate WA131M into the SERVICE position.
Moving into the SERVICE Position, p. 26
04.
As necessary, remov
e the sensor
(3)
.
Installing and Removing Sensors, p. 27
Note:
Do not tilt the coupling nut. Use a suitable spanning wrench (e.g., the one contained in
ZU0680 ser
vic
e set or ZU0740 ser
vice set).
05.
Using the spanning wr
ench
(1)
, loosen the c
oupling nut
(2)
counterclockwise.
06.
Pull the driv
e unit
(4)
out of the proc
ess unit
(5)
.
1)