5.3 Installing and Removing Sensors
5.3.1 Safety Instructions on Installing and Removing Sensors
WARNING! Process medium, possibly containing hazardous substances, can escape from the
SensoGate WA131M.
Follow the safety instructions.
CAUTION! Risk of cutting injuries from broken sensor glass.
Handle the sensor with care. Follow
the safety instructions in the related sensor documentation.
Note:
The outlet is used to drain trapped process medium and must not be closed. By moving the sen-
sor to the respective limit positions, pressurized process medium can enter the calibration chamber
and be compressed when the outlet is closed. This process medium may splash out during sensor re-
placement.
5.3.2 Solid-Electrolyte Sensor, Short Immersion Depth: Installation
1
2
3
10
9
3
7
8
6
5
4
01.
Move the SensoGate WA131M into the SERVICE position.
Moving into the SERVICE Position, p. 26
02.
Check outlet and leakage bores for escaping process medium. If process medium escapes, depres-
surize the process and perform troubleshooting.
03.
Check the compression r
ing
(4)
and O-r
ing
(5)
of the sensor
(2)
for correct positioning and
damage. Replace them if necessary.
04.
P
ush the sensor
(2)
into the SensoGate WA131M.
Note:
When tightening the sensor, the spring force of the “Immersion lock without a mounted solid-
electrolyte sensor” safeguard must be overcome.
05.
Tighten the sensor
(2)
using the spanning wr
ench
(10)
t
o max. 3 N
m (
A/F 19 mm). R
ecommended
tools: ZU0647 sensor spanning wrench
06.
Connec
t the cable bushing
(9)
t
o the sensor head
(1)
.