![Knick SensoGate WA131M Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/knick/sensogate-wa131m/sensogate-wa131m_user-manual_1980203007.webp)
SensoGate WA131M
7
The availability of safeguards is in part dependent on the version of SensoGate WA131M.
Environmental influences may affect the functionality of safeguards (e.g. components stuck together
by process medium).
1.4 Residual Risks
The product has been developed and manufactured in accordance with generally accepted safety
rules and regulations. However, it is not possible to rule out all risks.
1.4.1 Environmental Influences
The effects of moisture, ambient temperature, chemicals, and corrosion can negatively impact the safe
operation of the product. Please observe the following instructions:
• Only operate SensoGate WA131M in compliance with the stated operating conditions.
• If possible, install the product inside a protected area of the plant. Alternatively, take appropriate
measures to protect the SensoGate WA131M (e.g
. install ZU0759 pr
otective cap
1)
).
• If using aggressive chemical process media, adjust the inspection and maintenance intervals
accordingly.
• Adhering and sticky process media can impact the functionality of SensoGate WA131M (e.g
., b
y
causing components to stick together). Adjust the inspection and maintenance intervals accord-
ingly.
1.4.2 Accidental Loosening of the Process Connection
The coupling nut of the screw joint on process connections with a thread may become loose by acci-
dent. This may be caused by manual rotation of the rotating collar when moving to the limit positions
or by process-related vibrations.
Pressurized process medium may escape. Use of an appropriate retainer clamp or locking clamp is
strongly recommended.
Operating SensoGate WA131M without a retainer or locking clamp is at the risk of the operating com-
pany. The operating company must take action to rule out the possibility of the screw joint coupling
nut accidentally loosening.
1)
The ZU0759 protective cap protects against the effects of weather exposure and prevents the ingress of external
liquids or particles into the area of the sensor connections.
Содержание SensoGate WA131M
Страница 59: ...SensoGate WA131M 59 Appendix Return Form ...