SensoGate WA131M
21
2.7.2 Limit Switch
Note:
The limit switch is only available on versions of SensoGate WA131M with electronic limit signal.
The limit swit
ch
(1)
is a “simple apparatus” as defined in EN 60079-11 for use in explosive atmospheres
up to Z
The limit swit
ch
(1)
includes two reed switches (normally-open contacts), each of which is protected
b
y a 30 Ω ser
ies resistor.
Note:
Reed switches are sensitive to transient overruns of the limit values (e.g., due to cable capaci-
tance or inductance).
The limit swit
ch
(1)
has the following characteristics:
• Need not be marked according t
o EN 60079
• For connection to intrinsically safe circuits only
• Connection and ambient conditions:
◦
U
i
= 30
V
◦
I
i
= 100 mA
◦
P
i
= 750 mW
◦
C
i
= negligibly low
◦
L
i
= negligibly low
Temperature class
T6
T6
T5
T5
Equipment protection
level
Ga
Gb
Ga
Gb
Ambient temperature
range
-10 °C ... +45 °C
14 °F … 113 °F
-10 °C ... +60 °C
14 °F … 140 °F
-10 °C ... +57 °C
14 °F … 134.6 °F
-10 °C ... +70 °C
14 °F … 158 °F
• Isolation v
oltage: 500 V A
C between housing and terminals
• When installed, its stainless steel housing is grounded via the SensoGate WA131M.
• Verify the intrinsic safety before connecting the limit swit
ch
(1)
to an intrinsically safe circuit.
• M12 connector to EN 60947, 4-pole
4
3
1
2
1
1
4
3
30 Ω
2
30 Ω
M12
PROCESS
SERVICE
Содержание SensoGate WA131M