65
OK1, OK2 Inputs: Specify Level
Parameter setting/Module BASE/Inputs OK1, OK2
Note: HOLD mode (Setting: BASE module)
Menu Display
Setting the OK inputs
OK1 usage
• Open parameter setting
• Enter passcode
• Select “Module BASE”
• Select “Inputs OK1/OK2”
• Select “OK1 usage”
OK1/OK2 switching level
• Open parameter setting
• Enter passcode
• Select “Module BASE”
• Select “Inputs OK1/OK2”
• Specify active switching level
The BASE module provides 2 digital inputs
(OK1, OK2). The following functions (depending
on the parameter setting) can be started via a
control signal:
• OK1: “Off” or “HOLD”
• OK2: Select: System control / Function
control matrix. (“Off”, “Parameter set A/B”,
“Start KI recorder”)
You must specify the switching level
for the control signal:
(active 10...30 V or active < 2 V).
Module BASE
OK1
OK2
OK1
OK2
OK
Inputs OK1, OK2 (Administrator)
19.2°C
Return
OK2 usage
7.00 pH
OK1 usage
Off
HOLD
For OK2 usage see
“Function control matrix”
Inputs OK1, OK2 (Administrator)
19.2°C
Return
OK1 usage
7.00 pH
Input OK1
active 10 ... 30 V
active < 2 V
For OK2 usage see
“Function control matrix”
HOLD
Input OK2
OK
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...