137
Probe Maintenance via Protos 3400(X)
“Maintenance / Probe control” Menu
Maintenance
Return
25.6 °C
Module BASE 3400-021
Module PHU 3400-110
7.00 pH
Probe control
OK
Return
25.6 °C
Start probe maintenance
Single step control UNICAL
7.00 pH
OK
Unical 9000
Start program
-Please select-
Menu Display
Maintenance
Probe wear
Select “Probe control”.
The maintenance menu shows the
probe controller as a component of
the PHU 3400(X)-110 module.
Select using arrow keys,
confirm with enter
End Service
7.00 pH
25.6 °C
Probe maintenance
With the Unical “SERVICE” program,
the probe is moved into SERVICE
position. The individual program steps
are indicated in the display:
01: Probe in SERVICE
02: Cleaning agent
0020 sec
03: Rinse water ON
0060 sec
04: Rinse water OFF
0002 sec
05: Purge air ON
0005 sec
06: Purge air OFF
0002 sec
07: Wait position
08: Rinse water ON
0010 sec
09: Rinse water OFF
0002 sec
10: Probe in MEASURE
0005 sec
11: Program end
After end of servicing work the probe
moves back to “Measuring” position
(PROCESS).
Program: Service
Water pressure OK
3: Rinse water ON
Manual control
UNICAL
End Service
7.00 pH
25.6 °C
Program: Service
10: Probe in MEASURE
UNICAL
Maintenance of Unical 9000(X),
Uniclean 900(X)
Probe maintenance
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...