104
Display text
Time [s]
01: Probe in SERVICE
02: Cleaning agent
0020 sec
03: Rinse water ON
0060 sec
04: Rinse water OFF
0002 sec
05: Purge air ON
0010 sec
06: Purge air OFF
0002 sec
07: Wait position
> position will be held until next command (e.g., DCS)
08: Rinse water ON
0010 sec
09: Rinse water OFF
0002 sec
10: Probe in MEASURE
0005 sec
11: Program end
Parameter Setting: Program Flows
Parking
Program step
Probe: Measure
Probe: Service
Rinse water
Purge air
(AUX1 valve)
Cleaning agent
(Pump III)
HOLD
DCS ML3
(Program running)
01
02
03
04
05
06
07
08
09
10
Stroke 1 Stroke 2
Exposure time
Wait
t
[s]
Message (display)
W
ait p
osition
Parking is started via a DCS input signal at input BIN3
of the Uniclean 900 probe controller.