66
Switching Parameter Sets via OK2
Parameter setting / System control / Function control matrix
Note: HOLD mode (Setting: BASE module)
Parameter Sets
2 complete parameter sets (A, B) can be stored in the analyzer.
You can switch between the parameter sets using the OK2 input.
The currently activated set can be signaled by a relay contact.
An icon in the measurement display shows which parameter set is active:
or
Menu Display
Parameter sets
Signaling Active Parameter Set via
Relay Contact
• Open parameter setting
• BASE module
• Select contact
• Usage: “Parameter set ...”.
Selecting Parameter Set (A, B) via
OK2 Input
• Open parameter setting
• System control
• Function control matrix
• Select “OK2”
• Connect “Parameter set A/B”
Contact K3 (Administrator)
Abort
24.8°C
Contact type
OK
NAMUR maintenance!
Usage
ON delay
NAMUR HOLD
Limit value
Rinse contact
Parameter set B active
USP output
83.1 % Air
OFF delay
Note
The selection has no effect when working on SmartMedia card with
SW 3400-102.
24.8°C
Left softkey
83.1 % Air
Return
Connect
Function control matrix (Administrator)
Input OK2
Right softkey
Profibus DO 2
ParSet KI rec. Fav Unical
–
–
–
–
–
–
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...