139
Single Step Control
“Maintenance / Probe control” Menu
Return
25.6 °C
Start probe maintenance
Single step control UNICAL
7.00 pH
OK
Unical 9000
Start program
-Please select-
Menu Display
Maintenance
Probe wear
Single step control UNICAL
Each program can be executed in
single-step mode. A safety prompt is
displayed before the program starts.
The individual program steps are
indicated in the display. A "Cleaning"
program might be displayed as
follows:
01: Probe in SERVICE
02: Cleaner
0020 sec
03: Rinse water ON
0060 sec
04: Rinse water OFF
0002 sec
05: Probe in MEASURE
0005 sec
06: Program end
Manual control
Yes
25.6 °C
Start program
7.00 pH
No
Single step control UNICAL
Safety prompt:
Execute function?
-Please select-
End
01: Probe in SERVICE
02: Cleaner
0020 s
03: Rinse water ON
0060 s
25.6 °C
7.00 pH
Clean (Single step)
04: Rinse water OFF
0004 s
Program status: OK
Single step
Return
25.6 °C
Start probe maintenance
Single step control UNICAL
7.00 pH
OK
Unical 9000
Start program
-Please select-
Probe wear
Start program
Here you can select a program for
test purposes:
Clean
Cal 2point
Cal 1point
Parking
User 2
User 3
Manual control
Single step control
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...