95
System Forecast: Diagnostic Messages
Menu: Diagnostics / System forecast
Menu Display
System forecast: Diagnostics
Message list
When you have activated "Monitoring
On" in the System forecast menu, a
violation of a preset tolerance will
generate a message text. These texts
can be viewed in the message list.
System forecast
In operation, travel times are saved
for every cycle (max. 200) and are
compared with the active reference
profile.
If "System forecast" is selected in
the diagnostics menu, the active
reference profile and probe profiles
already recorded are displayed (this
may take some time). If tolerance lim-
its are exceeded in the probe travel
profile, the error number is displayed
after the probe profile.
NOTICE
In the Protos/Unical system, the last
200 travel cycles are recorded in vola-
tile memory, i.e. the data is lost when
the supply voltage is switched off.
Recording on a SmartMedia card is
therefore recommended – see follow-
ing page.
Return
25.6 °C
7.00 pH
12/21/16 12:37 S a06 b17 c25 d33 e02 CM
Load probe travel profile
12/21/16 12:37 P a09 b16 c26 d34 e02 CM User
System forecast
For better comparability, the travel
cycles can be displayed using the right
softkey:
- in Service “S“ and Process “P“
directions
- in Service "S" direction only
- in Process "P" direction only
S >>>
Return
25.6 °C
7.00 pH
System forecast
S >>>
Message list
U168
Return
UNICAL
active
1 messg.
SP SERVICE position not functioning
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...