134
Calibration / Adjustment
ISFET Zero Adjustment
ISFET Zero Adjustment
Notice:
The PHU 3400(X)-110 module does not supply the power for operating the
ISFET adapter.
When measuring with an ISFET sensor (Durafet, InPro 3300), the nominal zero
point must be adjusted each time a new sensor is connected (to adjust the
operating point). The adjustment for that sensor remains stored in the analyzer.
Afterwards, you should perform a two-point calibration using one of the
following methods:
• Calimatic: automatic calibration
• Manual entry of buffer values
• Data entry: premeasured electrodes
During calibration the module is in HOLD mode. Current outputs and relay
contacts of the module behave as configured (BASE module).
ISFET zero adjustment