129
Menu Display
Product calibration
Product calibration
Product calibration is performed
in 2 steps.
Prepare sampling,
Start with softkey or enter.
Step 1
Take sample.
Save measured value and temperature
at the moment of sampling (“Save”
softkey or enter)
Press meas to return to measurement.
Exception:
Sample value can be measured on the
site and be entered immediately.
To do so, press “Input” softkey.
Step 2
Lab value has been measured.
When you open the Product calibra-
tion menu again, the display shown
on the left appears:
Enter reference value ("Lab value").
Confirm with OK
or repeat calibration.
Product calibration
Return
Cal medium: Product
Cal by taking sample
and input of pH value
25.6 °C
7.00 pH
Product calibration
Step 1: Sampling
“Save” the sample value
“Input” lab value
Measured value
25.0°C
7.00 pH
+25.0°C
Temperature
7.00 pH
Input
Save
Product calibration
Abort
Step 2: Lab value
Input sample lab value
25.6 °C
7.00 pH
OK
+
7
.15 pH
Lab value
Start
Adjustment
Press “Adjust” to take over the values
determined during calibration for
calculating the measured variables.
25.6 °C
7.00 pH
Calibration data record
Cal mode
Response time
Slope
Product calibration
058.0 mV/pH
End
Adjust
0000 sec
Calibration
Zero
21.05.10 09:20
+07.00 pH
Содержание Protos 3400
Страница 12: ...12 ...
Страница 160: ...160 ...
Страница 177: ...177 Appendix Buffer Table Mettler Toledo ...
Страница 179: ...179 Appendix Buffer Table DIN 19267 ...
Страница 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Страница 182: ...182 Appendix Buffer Table Hamilton ...
Страница 183: ...183 Appendix Buffer Table Kraft ...
Страница 184: ...184 Appendix Buffer Table Hamilton A ...
Страница 185: ...185 Appendix Buffer Table Hamilton B ...
Страница 186: ...186 Appendix Buffer Table HACH ...
Страница 206: ...206 Index ...
Страница 207: ...207 Index ...