76
Display text
Time [s]
01: Rinse water ON
0010 s
02: Rinse water OFF
0002 s
03: Buffer I
0000 s
04: Cal Buffer 1
05: Rinse water ON
0010 s
06: Rinse water OFF
0002 s
07: Buffer II
0000 s
08: Cal Buffer 2
09: Rinse water ON
0010 s
10: Rinse water OFF
0002 s
11: Program end
Parameter Setting: Program Flows
Cal 2point (Short-Time), Media Monitoring Off
Program step
Probe: Measure
Probe: Service
Rinse water
HOLD
01
02
03
04
05
06
07
08
09
[t]
s
Message (display)
10
Buffer 4.01
(Pump I)
Buffer 7.00
(Pump II)
Calibration with
1st buffer
Calibration with
2nd buffer
Cal 2point (short-time) can also be started via a DCS input signal at input BIN2