12.4 Conditions for specification
Technical data has been determined in accordance with the ASTM stan-
dard E1657-98: “Standard Practice for Variable-Wavelength Photometric
Detectors Used in Liquid Chromatography”.
Reference conditions: test cell at wavelength 254 nm/8 nm with reference
wavelength 360 nm/30 nm, time constant 2 s, data rate 1 Hz.
Linearity (5%): Linearity is measured with caffeine at 274 nm/8 nm and
time constant 2 s with a flow cell path length 10 mm.
Performance tests should be done with a completely warmed up optical
unit (> 2 h). ASTM measurements require that the detector and/or lamp
should be turned on at least 24 h before start of testing. ASTM drift tests
require a temperature change below 2 °C/h over a 1 h period.
13. Reorders
This list for reorders is valid for the time the document has been pub-
lished. Deviations afterwards are possible.
For reorders of spare parts use the enclosed packing list. Contact the
Technical Support in case there are any questions on spare parts or
acces sories.
Further information on spare parts and accessories can be found online:
www.knauer.net
13.1 Devices and accessories
Name
Order No.
DAD 6.1L
Diode array detector DAD 6.1L with test cell
ADC11
DAD 2.1L
Diode array detector DAD 2.1L with test cell
ADC01
MWD 2.1L
Multi wavelength detector MWD 2.1L with test
cell
ADB01
Upgrade MWD 2.1L to DAD 2.1L
ADB01UMBAU
Instructions
Instructions English/German
V6700
Installation
Qualification
DAD 6.1L, DAD 2.1L, MWD 2.1L
VIQ_INST
Operation
Qualification
DAD 6.1L, DAD 2.1L, MWD 2.1L
VOQ_DAD
Lamps
High intensity deuterium lamp (for DAD 6.1L)
AZL01
Halogen lamp (for DAD 6.1L)
AZL02
Standard deuterium lamp (for DAD 2.1L and
MWD 2.1L)
A5193
Drainage system
Corrugated hose, 16 cm, PE grey
A9846-1
Funnel
P6431
Exhaust
P6432