Installation and initial startup
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
Connection
Function
Autozero
TTL Input
Low active
Secure switching threshold at least 10 mA
A signal (short-circuit to ground) sets the measu-
ring signal to zero.
+5 V
Provides a voltage of 5 V with respect to GND.
This makes it possible to supply a load that is
switched by an EVENT.
Protection: 5 V–50 mA
GND
Reference point of the voltage at the signal
inputs.
+24 V Valve
Event-controlled switching of 24 V
Protection against GND: 24 V–200 mA
Valve OUT
Output is active until the valve condition has
been eliminated.
GND
Reference point of the voltage at the signal
inputs.
4.8.8 Connecting the pin header
To control one device through another, the pin header is used. To use
remote control, you have to connect cables to the pin header. The single
ports are used to exchange control signals.
The device is switched off.
The power plug is disconnected from the device.
Depressor tool
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
cau ses a short circuit.
Turn off the device before connecting cables.
Pull the power plug.
Electronic defect
Electrostatic discharge can destroy the electronics.
Wear a protective bracelet against electrostatic discharge and ground.
Prerequisites
Tools