62
Technical data
AZURA® Detector DAD 6.1L, DAD 2.1L, MWD 2.1L Instructions, V6700
Communication
Interfaces
LAN (RJ-45), RS-232 (SUB-D 9), multi-pin
connector, analog (RCA cinch connector)
Control
Front panel, Mobile Control, software, event
con trol, analog, terminal protocol
Inputs
Error (IN), Start (IN), Autozero, Event 1–2 (TTL,
OC, Relay)
Outputs
Error (OUT), +5 V, Valve +24 V, Valve (OUT),
Start (OUT)
Analog output
Wavelength 0–10 V, flow rate 0–10 V
Analog output
1 x 0–5 V scalable, 20 bit, offset adjustable
Technical Parameters
GLP
Detailed report including lamp recognition,
ope rating hours, lamp operating hours, number
of lamp ignitions
Display
Mobile Control (optional)
Ambient conditions
Temperature range 4–40 °C, 39.2–104 °F
Humidity: below 90 %, non-condensing
General
Power supply
100–240 V, 50–60 Hz, 75 W
Dimensions
361 x 158 x 523 mm (H x W x D)
Weight
12.2 kg
Leak sensor
yes
12.3 MWD 2.1L
Detection
Detector type
Multiwavelength detector
Detection channels
8 (digital), 4 (analog)
Light source
Deuterium (D
2
) lamp with integrated GLP chip
Wavelength range
190–700 nm
Spectral bandwidth
< 8 nm at H
a
line (FWHM)
Note:
digital bandwidth 1–32 nm
Wavelength
accuracy
± 1 nm