Using RBW = 8 MHz to get accurate Gate settings
9.
Change the Res BW for
Ref Channel from 8
MHz to 100 kHz.
a. Press
Meas Setup
,
Ref
Channel
,
Res BW
, and enter
100 kHz
.
10.
Change the Res BW for
each Offset from 8 MHz
to 100 kHz.
a. Press
Meas Setup
,
Offset/Limits
,
Select Offset
,
Offset A
,
More
,
Res BW
and
enter
100 kHz
.
b. Set the same Res BW for Offset
B, C, D, E, F, respectively.
By default, Offset E and F are turned off.
11.
Turn on Average
function to smooth the
display, if necessary.
a. Press
Meas Setup
,
Avg/Hold
Num
, and enter the average
number for your test, like
10
.