Keysight WLAN Measurement Guide 85
Making WLAN Measurements
Troubleshooting WLAN Measurements
Troubleshooting WLAN Measurements
The following table provides some with descriptions you might encounter when
making WLAN measurements.
For more information on error messages see
Instrument Messages
.
Table 2-7
Error
Numbers
Messages
Notes
-221
Setting Conflict
The parameter setting is not available for current
standard.
-224
Illegal Parameter
The parameter setting is not available for current
settings.
215
Sync error; Preamble not found
The preamble code is not detected.
217
Demod error; Invalid packet type
Packet type detected is not recognized.
217
Demod error; Acq time too short
Not an entire Payload is captured, that is, the acquisition
time is too short.
159
Settings Alert; Parm/data mismatch
The actual captured data length is not long enough, or,
the offset from TxFrequency to center frequency is not an
integer multiple of 1 MHz.