Performance Verification and Calibration
This chapter contains performance verification and calibration procedures. The
performance verification procedures allow you to verify that the power supply is
operating within its published specifications. The calibration procedures show
how to make voltage and current adjustment on the power supply.
Keysight Technologies calibration services
When your instrument is due for calibration, contact your local Keysight Service
Center for a low-cost recalibration.
Calibration interval
One year interval is recommended for most applications. Accuracy specifications
are warranted only if calibration is made at regular calibration interval. Accuracy
specifications are not warranted beyond the one year calibration interval. Keysight
does not recommend extending calibration intervals beyond two years for any
application. Keysight recommends that complete re-adjustment should always be
performed at the calibration interval. This will increase your confidence that the
U8001A and U8002A will remain within specification for the next calibration
interval. The re-adjustment provides the best long-term stability and accuracy.
Содержание U8001A
Страница 8: ...8 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 69: ...Service Guide 5 Keysight U8001A U8002A User s and Service Guide 69 Figure 5 4 Graph of load transient response ...