5
Service Guide
54
Keysight U8001A/U8002A User’s and Service Guide
Line Voltage Conversion
Line voltage conversion is accomplished by adjusting two components: the line
voltage selection switch and the power-line fuse on the rear panel.
WARNING
Shock Hazard
. Operating personnel must not remove power supply covers.
Component replacement and internal adjustment must be made only by
qualified service personnel.
1
Remove the power cord and
slide the fuse cover upwards
2
Pull the fuse plunger to remove
the fuse from the fuse holder
3
Remove the voltage selector
PCB using a flat-blade
screwdriver.
4
Replace the correct fuse and
orientate the PCB based on the
correct voltage.
Содержание U8001A
Страница 8: ...8 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 12: ...12 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U8001A U8002A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 69: ...Service Guide 5 Keysight U8001A U8002A User s and Service Guide 69 Figure 5 4 Graph of load transient response ...