5 Adjust for the desired output current.
– Press on the “Voltage/Current” button. When seeing the current value, “A”
blinks, turn the knob to adjust for the desired output current value.
6 Return to the meter mode.
– Press on the “Display Limit” button to return to the meter mode. The LIMIT
annunciator will be turned off.
7 Enable the output.
– Press on the “Output On/Off” button to enable the output.
– The OFF annunciator turns off and constant current (CC) annunciator turns
on.
– The display is now in meter mode.
NOTE
Verify that the power supply is in constant current mode. Ensure that the
constant current (CC) annunciator is on. If the constant voltage (CV) annunciator
is on instead, choose a higher voltage limit.