To disable the OCP
1
Press on the “Over Current” button to disable the OCP. The OCP annunciator
will eventually turn off.
To clear the over current condition
1
OCP trip can only occur when the output is enabled. When the OCP condition
occurs:
– The output is disabled and the OFF annunciator will be turned on.
– If the keylock feature was enabled, it will be disabled. The LOCK annuciator
will be turned off.
– The OCP annunciator will fast-blink continuously.
– The display will show “TRIP”.
2
To clear the OCP trip conditon, press on the “Over Current” button again.
NOTE
If OCP trip persists, decrease the current limit settings to clear the tripping.