3 Calibration
68
Model N6705 Service Guide
Current Calibration
High Range Programming and Measurement - All Models
Do not connect anything to the Sense terminals.
Step 1. Connect a precision shunt resistor to the output. The shunt resistor
should be able to measure at least 120% of the output’s rated full-scale
current (see Cal Setup B). Connect the Keysight 3458A across the shunt
resistor.
Step 2. Select the full-scale current programming range. The value to program a
range must be the maximum current of the range. This example selects
the 8 A full-scale range of output 1.
For models
N6781A, N6782A, and N6784A, select the 3 A range - which
calibrates both the 3 A and 1 A ranges.
Front Panel:
SCPI Command:
From the Calibration menu, select
Current, High Range Prog and Meas.
Then select Start.
For N6785A, N6786A
CAL:CURR 8
, (@1)
For N6781A, N6782A, N6784A
CAL:CURR 3
, (@1)
Step 3. Select the first current calibration point.
Front Panel:
SCPI Command:
The first calibration point is automatically
selected “Enter P1 measured data” should be
displayed.
CAL:LEV P1
*OPC?
Step 4. Calculate the shunt current (I=V/R) and enter the data. This value may be
positive or negative.
Front Panel:
SCPI Command:
Enter the calculated value into the data field
and then select Next.
CAL:DATA <data>
Step 5. Select the second current calibration point. Wait a minimum of 5 minutes
for the internal temperature to stabilize.
Front Panel:
SCPI Command:
The second calibration point is automatically
selected “Wait 5 minutes, press Next” should
be displayed. Wait 5 minutes and select
Next. “Enter P2 measured data” should be
displayed.
CAL:LEV P2
*OPC?
Step 6. Calculate the shunt current (I=V/R) and enter the data.
Front Panel:
SCPI Command:
Enter the calculated value into the data field
and then select Next.
CAL:DATA <data>
Содержание N6705
Страница 1: ...Service Guide Keysight N6705 DC Power Analyzer ...
Страница 8: ......
Страница 104: ......
Страница 116: ......
Страница 120: ...6 Replaceable Parts 120 Model N6705 Service Guide Parts Location Diagrams Figure 6 1 External Parts ...
Страница 121: ...Replaceable Parts 6 Model N6705 Service Guide 121 Figure 6 2 Top Cover Removed ...
Страница 122: ...6 Replaceable Parts 122 Model N6705 Service Guide Figure 6 3 Top Chassis Removed ...
Страница 123: ...Replaceable Parts 6 Model N6705 Service Guide 123 Figure 6 4 Front Panel Figure 6 5 Handle Assembly ...
Страница 124: ......
Страница 132: ......