3 Calibration
66
Model N6705 Service Guide
Positive Voltage Limit - Models N6781A, N6782A, N6784A
Step 1. Select the positive voltage limit range. The value to program a range
must be the maximum voltage of the range to be calibrated. This
example selects the 20 V positive limit range of output 1.
Front Panel:
SCPI Command:
From the Calibration menu,
select Voltage, Positive Voltage
Limit. Then select Start. Select
the 20 V range.
CAL:VOLT:LIM:POS 20
, (@1)
Step 2. Select the first voltage calibration point.
Front Panel:
SCPI Command:
The first calibration point is automatically
selected. “Enter P1 measured data” should be
displayed.
CAL:LEV P1
*OPC?
Step 3. Measure the output voltage with the DMM and enter the data.
Front Panel:
SCPI Command:
Enter the value from the external DMM into
the data field and then select Next.
CAL:DATA <data>
Step 4. Select the second voltage calibration point.
Front Panel:
SCPI Command:
The second calibration point is automatically
selected. “Enter P2 measured data” should be
displayed.
CAL:LEV P2
*OPC?
Step 5. Measure the output voltage with the DMM and enter the data.
Front Panel:
SCPI Command:
Enter the value from the external DMM into the
data field and then select Next. When
calibration is complete, select Close.
CAL:DATA <data>
Step 6. Repeat steps 1 through 5 for the 6 V positive limit range.
Содержание N6705
Страница 1: ...Service Guide Keysight N6705 DC Power Analyzer ...
Страница 8: ......
Страница 104: ......
Страница 116: ......
Страница 120: ...6 Replaceable Parts 120 Model N6705 Service Guide Parts Location Diagrams Figure 6 1 External Parts ...
Страница 121: ...Replaceable Parts 6 Model N6705 Service Guide 121 Figure 6 2 Top Cover Removed ...
Страница 122: ...6 Replaceable Parts 122 Model N6705 Service Guide Figure 6 3 Top Chassis Removed ...
Страница 123: ...Replaceable Parts 6 Model N6705 Service Guide 123 Figure 6 4 Front Panel Figure 6 5 Handle Assembly ...
Страница 124: ......
Страница 132: ......