Model N6705 Service Guide
45
Calibration Set-up
The following figures show the calibration test set-ups. Connect all leads
to the Output terminals as shown in the following figures. Twist the + and
- load leads to reduce noise and inductance.
+
−
1
+
−
2
AUX VOLTAGE
MEASUREMENT
+
E.
C.
+
A.
Keysight 3458A
configured as a
voltmeter
Disable 4-wire
sense. LED
indicator is off.
B.
Current
shunt
+
Keysight 3458A
configured as a
voltmeter
Disable 4-wire
sense. LED
indicator is off.
Keysight 3458A
configured as a
voltmeter
Enable 4-wire sense.
Press [Settings], then
select Advanced.
Select 4 wire.
+
D.
Keysight 3458A
configured as
an ammeter
Disable 4-wire
sense. LED
indicator is off.
Electronic load
+
+
F.
Disable 4-wire
sense. LED
indicator is off.
Keysight 3458A
configured as a
voltmeter
Disable 4-wire
sense. LED
indicator is off.
External
DC supply
Current
monitor
+
+
G.
Disable 4-wire
sense. LED
indicator is off.
Keysight 3458A
configured as a
voltmeter
Содержание N6705
Страница 1: ...Service Guide Keysight N6705 DC Power Analyzer ...
Страница 8: ......
Страница 104: ......
Страница 116: ......
Страница 120: ...6 Replaceable Parts 120 Model N6705 Service Guide Parts Location Diagrams Figure 6 1 External Parts ...
Страница 121: ...Replaceable Parts 6 Model N6705 Service Guide 121 Figure 6 2 Top Cover Removed ...
Страница 122: ...6 Replaceable Parts 122 Model N6705 Service Guide Figure 6 3 Top Chassis Removed ...
Страница 123: ...Replaceable Parts 6 Model N6705 Service Guide 123 Figure 6 4 Front Panel Figure 6 5 Handle Assembly ...
Страница 124: ......
Страница 132: ......