1
Introduction
40
Keysight N1913/1914A User’s Guide
Rear Panel Connections
7
8
6
9
4
3
2
1
5
No.
Connections
1
VGA Output (Option 201)
2
Ground Connector
3
USB Type A port
4
Recorder 1/2
Recorder output (two outputs are fitted to dual channel meters) connections are made via BNC connectors. This
output produces a DC voltage that corresponds to the power level of the channel input.
5
AC Inlet
This power meter has an auto configuring power supply. This allows it to operate over a range of voltages without
manually being set to a certain voltage.
6
Trig In/Trig Out
Trigger input and output connections are made via BNC connectors.
7
USB Mini-B port
This USB port is used only for remote interface connection.
8
LAN
9
GPIB
This connector allows the power meter to be controlled remotely using the General Purpose Interface Bus.