Using 8480 Series Power Sensors
5
Keysight N1913/1914A User’s Guide
157
8480 Series Sensors Connection Requirements
Table 5-1
8480 Series connection requirements
Sensor
Connection Requirements
8481A
8481H
8482A
8482H
These power sensors connect directly to the POWER REF.
8481D
8484A
Prior to calibration, a Keysight 11708A 30 dB reference attenuator should be connected
between the power sensor and the POWER REF. Remove this attenuator from the power
sensor input before making measurements.
8483A
This power sensor requires a 75
Ω
(f) to 50
Ω
(m) N-Type adapter (1250-0597) to connect
to the POWER REF. Remove this adapter before making measurements.
R8486A
Q8486A
V8486A
W8486A
E8486A
R8486D
Q8486D
These waveguide power sensors have two connectors. Use the N-Type connector to
calibrate the power meter.
8481B
8482B
These power sensors are configured with an attenuator. Prior to calibration this attenuator
must be removed. Replace the attenuator before making measurements.
8485A
This power sensor requires an APC 3.5 (f) to 50
Ω
(m) N-Type adapter (08485-60005) to
connect to the POWER REF. Remove this adapter before making measurements.
8485D
Prior to calibration, a Keysight 11708A 30 dB reference attenuator and an APC 3.5 (f) to
50
Ω
(m) N-Type adapter (08485-60005) should be connected between the power sensor
and the POWER REF. Remove this attenuator and adaptor before making measurements.
8487A
This sensor requires an APC 2.4 (f) to 50
Ω
(m) N-Type adapter (08487-60001) to connect
to the POWER REF. Remove this adapter before making measurements.
8487D
Prior to calibration, a Keysight 11708A 30 dB reference attenuator and an APC 2.4 (f) to
50
Ω
(m) N-Type adapter (08487-60001) should be connected between the power sensor
and the POWER REF. Remove this adapter before making measurements.
Содержание 1914A
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters User s Guide...
Страница 24: ...24 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 54: ...1 Introduction 54 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 206: ...7 Using U2000 Series USB Power Sensors 206 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 240: ...9 Maintenance 240 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 242: ...10 Characteristics and Specifications 242 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...