Using N8480 Series Power Sensors
6
Keysight N1913/1914A User’s Guide
189
3
Press
to complete the selection of the calibration factor table.
4
Press
again to display the measurement screen.
shows which
offset table is selected.
Figure 6-9
Frequency dependent offset indicator
5
To change the frequency, press
and use the
and
keys to
highlight the
Frequency
field.
6
Press
to display the
Frequency
pop-up window. Use the numeric
keypad to enter the required value in the
Frequency
pop-up window.
7
To confirm your choice, press the appropriate unit softkey.
8
Connect the power sensor to the signal to be measured.
9
The corrected measurement result is now displayed.
NOTE
When no data is contained in the highlighted table, the
key is disabled
(grayed out).
Table
Done
Done
Sensor Table 2 Selected
NOTE
If the measurement frequency does not correspond directly to a frequency in the
sensor calibration table, the power meter calculates the calibration factor using
linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor
calibration table, the power meter uses the highest or lowest frequency point in
the sensor calibration table to set the calibration factor.
Содержание 1914A
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters User s Guide...
Страница 24: ...24 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 54: ...1 Introduction 54 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 206: ...7 Using U2000 Series USB Power Sensors 206 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 240: ...9 Maintenance 240 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 242: ...10 Characteristics and Specifications 242 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...