6
Using N8480 Series Power Sensors
192
Keysight N1913/1914A User’s Guide
There are also ten sensor calibration tables named
CUSTOM_0
through
CUSTOM_9
. These tables do not contain any data when the power meter is
shipped from the factory.
Editing frequency dependent offset tables requires the following steps:
1
Identify and select the table you want to edit
2
Rename the table
3
Enter the frequency and offset pairs
4
Save the table
Procedure
First, select the table you want to edit as follows:
1
Press
,
,
to display the
Sensor Tbls
screen.
Figure 6-11
“Sensor Tbls” screen
2
Choose the table you want to edit using the
and
keys. Press
to display the
Edit Cal
screen as shown in
.
Tables Sensor Cal Tables
Edit Table