Using N8480 Series Power Sensors
6
Keysight N1913/1914A User’s Guide
177
Power Meter Configuration Changes
The N1913/1914A EPM Series power meter recognizes when a N8480 Series
power sensor is connected. The N8480 Series power sensors’ (excluding Option
CFT) calibration data is automatically read by the power meter. In addition, the
auto-averaging settings shown in
are automatically configured.
Figure 6-1
Auto-averaging settings
Sensor Dynamic
Rang
e
Nu
mber of
Averag
es
Upper Rang
e
Lower Rang
e
Sensor Power
Minimum
1
2
3
4
1
1
2
8
Resolution Setting
2
2
4
32
2
2
4
32
2
2
16
256
2
8
128
128
Maximum
4
64
256
512
Sensor Power
0 dBm
–1 dBm
–30 dBm
–10 dBm
–20 dBm
N8481/2/5A
NOTE
These values are valid only for the power meter channel connected to a N8480
Series power sensor. Averaging settings can also be manually configured. Refer
“Setting Measurement Averaging”
on page 73 for more details.
Содержание 1914A
Страница 1: ...Keysight N1913 1914A EPM Series Power Meters User s Guide...
Страница 24: ...24 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 54: ...1 Introduction 54 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 206: ...7 Using U2000 Series USB Power Sensors 206 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 240: ...9 Maintenance 240 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 242: ...10 Characteristics and Specifications 242 Keysight N1913 1914A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...