Making Measurements
Making an Average Power Measurement
38
Setting the Power Meter’s Upper and Lower
Limits
The internal Power Meter can be configured to detect
when a measurement has failed a user predefined upper
and lower limits.
1 Press
[Preset]
to perform a factory preset.
2 Press
[MODE]
>
{Power Meter} > [ENTER]
to turn on
the power meter.
3 Press
{Meas Setup}
>
{Limits}
to access the limits
menu.
4 Press
{Limits}
to activate the limits function. Each time
the softkey is pressed, the selected option changes.
5 Press
{Upper Limits}
and enter the high limit value
using the numeric keypad. Then press
{dBm}
to
complete the setup.
6 Press
{Lower Limits}
and enter the low limit value
using the numeric keypad. Then press
{dBm}
to
complete the setup.