Key Reference
System
85
Upgrade
Enables the firmware upgrade process.
1 Copy the upgrade application onto the root level of a
USB memory stick.
2 Insert the USB memory stick into the USB
interface of the tester on its top panel.
3 Press
[SYS] > {Upgrade}
to initiate firmware upgrade.
4 The tester will promotes a message to indicate
firmware upgrading is in process.
5 After upgrading completes, power off the tester then
reboot it.
Key access:
[SYS] > {Upgrade}
Status
Shows the systematic information of the tester.
Key access:
[SYS] > {Status}
Self Test
Runs a self test on temperature and hardware.
Key access:
[SYS] > {Self Test}
E Calib Test
Verifies the communication between the electronic
calibrator and the tester.
Key access:
[SYS] > {E Calib Test}
Intelligent
Toggles the ability to identify the calibration loads
(Open/Short/Load) from each other between on and off.
Key access:
[SYS] > {Intelligent}