Making Measurements
Making an Average Power Measurement
36
To Make a Basic Average Power
Measurement
You can follow the steps below to make a basic average
power measurement.
1 Press
[Preset]
to perform a factory preset.
2 Press
[MODE]
>
{Power Meter} > [ENTER]
to turn on
the power meter.
3 Zero and calibrate the meter. Press
{Zero}
to make a
Zero operation of the power sensor followed by a
calibration operation.
4 Connect the power sensor to the power Ref 50 MHz
port. The analyzer supports the U2000 Series power
sensors.
5 Connect the external attenuation, if required, with the
power sensor used.
6 Connect the analyzer to the base station. Connect the
power sensor to the signal to be measured. A typical
BTS connection is after the output of the power
amplifier and duplexer. The connection requires a 2 ft.
cable and two Type-N barrel connectors, in addition
to the power sensor and external attenuator.
Figure 3-1 Connection with base station