Using Functions
Some Helpful Hints
53
Repeatability
If two identical measurements are made with the tester,
the differences in the measured data should be so small
that they do not affect the value of the measurement.
Connector repeatability may be a cause of poor
measurement results. Connector repeatability problems
may be affected by:
• Dirty or damaged connectors
• Connections that have been made without using
proper torque techniques (this applies when
connections to the tester have been removed, then
reconnected).
Extending the life of your cables and
connectors
• Minimize the bending of test cables. A single sharp
bend can damage a RF cable.
• Avoid repeated connection and disconnection of all
RF cable connections.
• Do not bend cables at or near the RF connectors. This
may affect the RF performance.
• Inspect all connectors for dirt, nicks, and other signs
of damage or wear before connection. A bad
connector can ruin a good connector instantly.
• Clean dirty connectors. Dirt and foreign matter can
cause poor electrical connections, affect RF
measurements, and may permanently damage the
connectors.
Proper Connector Torque
CAU-
CAUTION
Never exceed the recommended connector torque values
when attaching cables or damage may result.
Connector
Torque
cm-kg
Torque
N-cm
Torque
in-lbs
Type-N
52
508
45
3.5 mm
9.2
90
8
SMA
5.7
56
5
Содержание N9330B
Страница 7: ...1 Overview 1 Overview...
Страница 13: ...7 Getting Started 2 Getting Started...
Страница 29: ...23 Making Measurements 3 Making Measurements...
Страница 45: ...39 Using Functions 4 Using Functions...
Страница 67: ...61 Key Reference 5 Key Reference...
Страница 93: ...87 Instrument Messages 6 Instrument Messages...
Страница 97: ...91 Menu Map 7 Menu Map...
Страница 98: ...Menu Map FREQ DIST 92 FREQ DIST FREQ DIST Start Home Signal Standard Stop End Page down Page up...
Страница 99: ...Menu Map AMPTD 93 AMPTD AMPTD Top Bottom...
Страница 104: ...Menu Map Marker 98...