Making Measurements
Calibrate the N9330B before making measurements
28
3 Press
[ENTER]
to start calibration.
Calibrate with a T-combo calibrator
Using a T-combo calibrator is a three step process
requiring the user to manually change the
open-short-load components. The default order of the
calibration process is open-short-load.
1 Press
[CAL]
to bring up the dialog box for calibration
guidance.
2 Connect the “Open” component of the T-combo
calibrator to the RF OUT connector of the tester.
Press
[ENTER]
to start the open calibration.
3 Connect the “Short” component of the T-combo
calibrator to the RF OUT connector of the tester.
Press
[ENTER]
to start the short calibration.
4 Connect the “Load” component of the T-combo
calibrator to the RF OUT connector of the tester.
Press
[ENTER]
to start the load calibration.
A “Calibrated” indicator will be displayed on the upper
left corner of the screen when the calibration is
completed.
Using User Cal Mode
A
User Cal Mode
is an optional
selection when using a mechanical calibrator. The User
Cal Mode allows the user to choose one cal standard
from the open-short-load to perform a customized
calibration.
For example, pressing
[CAL] > {User Cal Mode} > {Open}
initiates a calibration
using the Open calibration standard only.
NOTE
Never connect an electronic calibrator to the tester until
after
powering on the N9330B.
Содержание N9330B
Страница 7: ...1 Overview 1 Overview...
Страница 13: ...7 Getting Started 2 Getting Started...
Страница 29: ...23 Making Measurements 3 Making Measurements...
Страница 45: ...39 Using Functions 4 Using Functions...
Страница 67: ...61 Key Reference 5 Key Reference...
Страница 93: ...87 Instrument Messages 6 Instrument Messages...
Страница 97: ...91 Menu Map 7 Menu Map...
Страница 98: ...Menu Map FREQ DIST 92 FREQ DIST FREQ DIST Start Home Signal Standard Stop End Page down Page up...
Страница 99: ...Menu Map AMPTD 93 AMPTD AMPTD Top Bottom...
Страница 104: ...Menu Map Marker 98...